Shanghai, China · Fine blanking steel & precision strip specialist sales@fineblankingmachine.com · WhatsApp: 008613062870081
Home / Knowledge Hub / Stainless Steel / Medical Devices Guide

The Ultimate Guide to Stainless Steel Strip for Medical Devices

From grade selection to sterilization resistance to failure mode prevention — the complete guide to stainless steel strip for surgical instruments, implants and medical device components. Five grades compared, three sterilization methods tested, four surface treatments mapped, and the full manufacturing process flow.

Published Sep 10, 2026 Reading time ~16 min Reviewed by HS-FINEB Engineering Team
At a glanceMedical device stainless steel falls into two families: martensitic grades (4Cr13, SUS420J2) that harden by quenching to HRC 48–53 for cutting edges, and austenitic grades (SUS301, SUS304, SUS316L) that cannot be through-hardened but offer superior corrosion resistance and biocompatibility. The grade decision starts with the function: cutting instrument → martensitic; implant or corrosion-critical component → SUS316L; spring or clip inside a device → SUS301 (work-hardened); general housing or tray → SUS304. This guide compares all five grades on chemistry, hardness, corrosion resistance and application, then maps sterilization cycle performance (134°C steam, ethylene oxide, gamma radiation), surface treatments (passivation, electropolishing, bead blasting, PVD titanium), the seven-step manufacturing process, and the four common failure modes with prevention measures.

1. Why Stainless Steel for Medical Devices

Stainless steel dominates medical device manufacturing for three reasons that no other structural material matches simultaneously: it can be sterilized repeatedly without degrading, it can be manufactured to extremely tight tolerances and sharp edges, and it has a decades-long track record of biocompatibility and regulatory acceptance. From disposable surgical blades to reusable orthopedic implants, stainless steel strip is the starting material for a large share of the instruments that enter a hospital every day.

The challenge for a device engineer is that "stainless steel" is not one material — it is a family of alloys with widely different properties. A surgical blade and a surgical tray are both made of stainless steel, but they require opposite properties: the blade needs maximum hardness and edge sharpness, while the tray needs maximum corrosion resistance and cannot rust after a thousand autoclave cycles. Specifying the wrong grade is one of the most common root causes of medical device field failures.

1.1 The two families: martensitic vs austenitic

Medical stainless steels divide into two metallurgical families, and the division is absolute — you cannot have the properties of both in one grade.

Martensitic stainless steels (4Cr13, SUS420J2, and the broader 420/440 series) contain 12–14% chromium and 0.26–0.45% carbon. The carbon allows them to be austenitized at high temperature (980–1050°C) and quenched to form hard martensite, reaching HRC 48–53. This is what makes a surgical scalpel hold a sharp edge. The tradeoff is that martensitic steels are less corrosion-resistant than austenitic grades — the carbon combines with chromium to form chromium carbides at the grain boundaries, reducing the chromium available for the passive oxide layer. They will stain or pit if not properly passivated and dried after sterilization.

Austenitic stainless steels (SUS301, SUS304, SUS316L) contain 16–20% chromium and 6–15% nickel, with carbon held low (≤0.08%, or ≤0.03% for the L-grades). The nickel stabilizes the austenite phase at room temperature, which means these steels cannot be hardened by quenching — they are soft and ductile in the as-supplied condition. Their advantage is corrosion resistance: the high chromium content plus low carbon produces a robust passive chromium oxide layer that resists rust, body fluids and repeated sterilization. SUS316L adds 2–3% molybdenum for even better pitting resistance, making it the grade of choice for implants and instruments exposed to chloride-containing environments.

1.2 Regulatory and material traceability context

Medical device materials are subject to regulatory traceability requirements that go beyond ordinary industrial steel. In the US, the FDA expects device manufacturers to demonstrate that the material meets a recognized standard (such as ASTM F899 for surgical instruments, ASTM F138/F139 for surgical implants, or ISO 7153-1 for surgical instruments). In the EU, the MDR requires material traceability and compliance with harmonized standards. This means the mill test certificate (MTC) must report the exact chemistry, and the material must be traceable from coil to finished device through every processing step.

For strip suppliers, this translates to three requirements: full chemistry certification per heat, consistent mechanical properties within the specified band, and a clean surface free of embedded contaminants that could compromise passivation. Our mill test certificate guide explains what to check on an MTC for medical-grade material.

2. Medical Grade Stainless Steel Comparison

The table below compares the five most common stainless steel strip grades used in medical device manufacturing. Values are typical of the latest standard revisions; always confirm against the specific standard cited on your device specification and the supplier's mill test certificate.

GradeStandardC (%)Cr (%)Ni (%)Hardness (max)Corrosion resistancePrimary medical use
4Cr13GB/T 12200.36 – 0.4512.0 – 14.0—HRC 48 – 53 (quenched)Moderate (martensitic)Surgical blades, scissors, osteotomes
SUS420J2JIS G4304 / ASTM F8990.26 – 0.4012.0 – 14.0—HRC 48 – 53 (quenched)Moderate (martensitic)Surgical instruments, cutting tools, dental scalers
SUS301JIS G4305 / ASTM A666≤ 0.1516.0 – 18.06.0 – 8.0HV 430 – 520 (full-hard, cold worked)Good (austenitic, work-hardened)Device springs, clips, staples, snap-fit components
SUS304JIS G4305 / ASTM A240≤ 0.0818.0 – 20.08.0 – 10.5HV ≤ 200 (annealed)Very good (austenitic)Instrument housings, trays, tubing, non-implant components
SUS316LJIS G4305 / ASTM F138/F139≤ 0.0316.0 – 18.012.0 – 15.0HV ≤ 200 (annealed)Excellent (austenitic, Mo-alloyed)Implants, orthopedic devices, surgical instruments, corrosion-critical parts

Chemistry limits are indicative and based on standard specifications. 4Cr13 and SUS420J2 hardness values are after quenching (oil or air) from 980–1050°C and tempering at 150–300°C. SUS301 full-hard hardness is after cold rolling reduction of ~40–50%; annealed SUS301 is HV ≤ 200. SUS316L for surgical implants should comply with ASTM F138 (forged) or F139 (wire) — confirm the specific grade and standard at RFQ. Corrosion resistance ratings are relative and assume proper passivation; actual performance depends on surface condition and environment.

Three points from the table drive the grade decision:

  • Only 4Cr13 and SUS420J2 can produce a hard cutting edge. If the device cuts, scrapes or needs a wear-resistant surface, one of these two martensitic grades is required. SUS301 can be work-hardened to high strength but does not produce a sharp, durable edge like a quenched martensitic steel.
  • SUS316L is the only implant-grade option in this group. Its low carbon (≤0.03%) plus molybdenum (2–3%) gives it the best pitting resistance and the longest track record for permanent implants. SUS304 is used for non-implant instruments and housings; it is not recommended for permanent implantation due to lower corrosion resistance.
  • SUS301 is the spring grade. It sits between the two families — austenitic in the annealed condition but capable of extreme work hardening (up to HV 520 in full-hard condition). This makes it the standard for internal device springs, retaining clips and staples where high strength plus good corrosion resistance is needed without heat treatment.

The 4Cr13 stainless steel strip product page and SUS301 stainless steel strip product page cover these grades in detail with available thickness and width ranges.

3. Sterilization Cycle Corrosion Resistance

Every reusable medical device must survive repeated sterilization. The three dominant methods — steam autoclaving, ethylene oxide (EtO) and gamma radiation — impose very different stresses on the material, and the grade's performance varies accordingly. The table below summarizes the relative corrosion resistance of each grade under each sterilization method, based on typical industry practice and accelerated aging test data.

Sterilization methodConditions4Cr13 / SUS420J2SUS301SUS304SUS316L
Steam autoclave134°C, 2.1 bar, 4–18 min, moist△ ≥1,000 cycles if passivated & dried; risk of staining if surface iron present✓ ≥2,000 cycles✓ ≥5,000 cycles✓ ≥10,000 cycles
Ethylene oxide (EtO)37–63°C, dry gas, 1–6 hrs✓ No significant corrosion✓ No significant corrosion✓ No significant corrosion✓ No significant corrosion
Gamma radiation25–50 kGy, ambient, dry✓ Minimal effect; possible slight discoloration✓ Minimal effect✓ Minimal effect; possible slight discoloration✓ Minimal effect

Cycle counts are typical/indicative for properly passivated, clean surfaces with distilled or deionized water in the autoclave. Actual cycle life depends on surface finish (electropolished > mechanically polished > ground), water chloride content, drying practice and storage conditions. Martensitic grades are more sensitive to residual moisture and chloride — instruments should be dried immediately after each autoclave cycle. EtO and gamma sterilization operate in dry conditions and do not promote aqueous corrosion. Discoloration from gamma radiation is cosmetic and does not affect corrosion resistance or mechanical properties.

Steam autoclaving at 134°C is the most demanding sterilization method for stainless steel because it combines high temperature, high pressure and moisture — the three conditions that accelerate electrochemical corrosion. The risk is not uniform across grades: austenitic grades (SUS304, SUS316L) have enough chromium to maintain a stable passive layer even under hot, moist conditions, while martensitic grades (4Cr13, SUS420J2) have less free chromium (some is tied up in carbides) and are more vulnerable to staining and pitting if the surface is contaminated with free iron from grinding or handling.

The autoclave rule for martensitic instruments. A properly passivated, electropolished SUS420J2 instrument should survive at least 1,000 autoclave cycles without rust spots. If your instruments show rust after fewer than 100 cycles, the problem is almost never the grade — it is either (a) embedded free iron from grinding or contact with carbon steel tools, (b) incomplete or degraded passivation, or (c) chloride in the autoclave water. Test the water, re-passivate the instruments, and isolate stainless steel from carbon steel in the manufacturing workflow.

4. Surface Treatments for Medical Devices

The surface of a medical device is where corrosion starts, where bacteria adhere, and where the patient's tissue makes contact. Surface treatment is therefore not a cosmetic afterthought — it is a critical design parameter that affects corrosion resistance, cleanability, biocompatibility and wear life. The table below compares the four most common surface treatments applied to medical stainless steel.

TreatmentProcessSurface roughness Ra (μm)Corrosion resistanceBiocompatibilityTypical use
PassivationNitric or citric acid bath, removes free iron, restores Cr oxide layerUnchanged (0.2 – 0.8)Improved (removes corrosion sites)Excellent (standard for all medical SS)All medical stainless steel devices — baseline treatment
ElectropolishingElectrochemical anodic dissolution in acid bath, removes surface layer≤ 0.2 (typ. 0.1 – 0.2)Significantly improved (smoother + cleaner surface)Excellent (smooth surface resists biofilm)Implants, surgical instruments, reusable devices
Bead blastingFine glass or ceramic beads at controlled pressure, matte finish0.8 – 2.0Reduced if not passivated afterward (rough surface traps contaminants)Good (if passivated after blasting)Instrument handles, non-critical surfaces, aesthetic matte finish
PVD titanium coating (TiN)Physical vapor deposition of titanium nitride, ~2–5 μm layerUnchanged (follows substrate)Improved (barrier layer, but pinhole risk)Excellent (TiN is biocompatible, gold color)Cutting edges, wear surfaces, premium instrument lines

Ra values are typical; actual results depend on the starting surface and process parameters. Passivation per ASTM A967 (citric or nitric acid) is the baseline requirement for all medical stainless steel. Electropolishing per ASTM B912 is specified for implants and high-corrosion-risk instruments. Bead blasting must always be followed by passivation to remove embedded bead material and restore the passive layer. PVD TiN coating hardness is typically HV 1800–2400; coating adhesion depends on substrate preparation and is verified by scratch testing. For regulated devices, the surface treatment must be validated and documented in the device master record.

Passivation is the non-negotiable baseline — every medical stainless steel device should be passivated after manufacturing to remove free iron embedded during grinding, stamping and handling. Electropolishing goes a step further by dissolving the smeared surface layer and peaks, producing a smoother, more uniform surface that is easier to clean and more resistant to pitting. Bead blasting produces a matte finish for grip or aesthetics but creates micro-crevices that can trap contaminants — it must always be followed by passivation. PVD titanium nitride (TiN) coating adds a hard, gold-colored wear layer to cutting edges and high-wear surfaces, but it requires a defect-free substrate because any pinhole in the coating can initiate galvanic corrosion.

For a broader treatment of stainless steel surface finishes including BA, 2B and 2D mill finishes, see our steel surface finish guide.

5. Manufacturing Process Flow

Medical device stainless steel components follow a controlled process flow where each step affects the final properties and must be documented for regulatory traceability. The table below outlines the seven-step flow for a typical martensitic surgical instrument (the most complex route); austenitic components skip the heat treatment step.

StepProcessKey parametersWhat it controls
1Stamping / blankingPrecision die, slit edge burr ≤0.02 mm, lubricant compatible with later cleaningPart geometry, edge quality, starting surface
2Heat treatment (martensitic only)Austenitize 980–1050°C, quench (oil/air), temper 150–300°C; vacuum or controlled atmosphere to avoid decarb and scalingHardness (HRC 48–53), toughness, edge retention, dimensional stability
3GrindingSurface or profile grinding, coolant (chloride-free), sequential grit reductionFinal dimensions, edge geometry, surface roughness
4PolishingMechanical polishing (belt / buff) or electropolishing; Ra target per device specSurface smoothness, cleanability, corrosion resistance
5PassivationCitric acid (4–10% @ 40–70°C) or nitric acid (20–50% @ ambient); ASTM A967Removal of free iron, restoration of passive Cr oxide layer
6CleaningUltrasonic cleaning in validated detergent + DI water rinse; residue testing per ASTM F2459Removal of manufacturing residues (oil, grit, acid), bioburden control
7PackagingSterile barrier system (pouch or tray + lid), validated for sterilization methodSterility maintenance, protection from contamination and damage

Process parameters are typical for surgical instrument manufacturing. Exact parameters depend on the grade, part geometry and device specification. Heat treatment for martensitic grades must be performed in a vacuum or controlled-atmosphere furnace to avoid decarburization and oxidation — decarb on the cutting edge reduces hardness and edge life. Cleaning validation (per ASTM F2459 / ISO 19227) is a regulatory requirement for implantable and reusable devices. The medical surgical blade application page covers the blade-specific process in detail.

5.1 Stamping and forming

The stamping step sets the part geometry and the edge condition that will carry through every subsequent operation. For medical instruments, the slit edge burr from the coil must be controlled to ≤0.02 mm — a larger burr can become a stress riser after heat treatment or a site for corrosion initiation. The stamping lubricant must be compatible with the downstream cleaning process; lubricants that leave a silicone or heavy-oil residue can interfere with passivation and are not acceptable for medical devices. For thin-gauge components (0.10–0.50 mm), precision slitting with controlled knife clearance is essential — our edge burr guide covers the slitting parameters that control burr height.

5.2 Heat treatment for martensitic grades

Heat treatment is the step that transforms a soft, formable 4Cr13 or SUS420J2 blank into a hard, sharp instrument. The part is heated to 980–1050°C to form austenite, quenched (in oil or pressurized gas for vacuum furnaces) to form martensite, and then tempered at 150–300°C to relieve quenching stresses and set the final hardness. The furnace atmosphere is critical — vacuum or controlled-atmosphere furnaces prevent decarburization (loss of carbon from the surface) and oxidation. A decarburized surface layer will not harden properly, resulting in a soft edge that dulls quickly. For thin strip components (under 1 mm), the heat treatment distortion must be controlled by fixturing or press quenching to maintain flatness.

5.3 Grinding, polishing and passivation

After heat treatment, the part is ground to final dimensions and edge geometry. Grinding must use chloride-free coolant because chloride residues can cause stress corrosion cracking in martensitic stainless steel. The grit sequence progresses from coarse (shape the edge) to fine (set the surface), and each step must remove the damaged layer from the previous step. Polishing follows — mechanical polishing for standard instruments, electropolishing for implants and high-corrosion-risk devices. The final step before cleaning is passivation, which removes any free iron embedded during grinding and polishing and restores the chromium-rich passive layer. Passivation is validated by salt spray testing (ASTM B117) or copper sulfate testing (ASTM A967) to confirm the surface is free of active iron.

Precision stainless steel strip coils ready for medical device manufacturing

6. Common Failure Modes and Prevention

Medical device stainless steel failures follow predictable patterns. The table below lists the four most common failure modes, their root causes, and the prevention measures that address them at the material and process level.

Failure modeAppearanceRoot causePrevention
Pitting corrosionSmall, deep holes or rust spots on surface, often near edges or crevicesChloride exposure + broken passive layer; embedded free iron; incomplete passivationUse SUS316L for chloride environments; ensure complete passivation (ASTM A967); electropolish for critical surfaces; use DI water in autoclave; isolate from carbon steel
Stress corrosion cracking (SCC)Fine cracks perpendicular to tensile stress, often branching; appears after months/yearsCombination of tensile residual stress + chloride + susceptible microstructure (martensitic)Stress relief after forming/grinding; use low-carbon austenitic grades (316L) for SCC-prone applications; avoid chloride contact; control grinding heat to prevent residual stress
Fatigue failureSmooth fracture surface with beach marks; initiates at surface defect or edgeCyclic loading above fatigue limit; surface defects (grinding marks, burrs, pits) act as initiation sitesElectropolish to remove surface defects; control edge burr ≤0.02 mm; design within fatigue limit (SUS301 full-hard for springs); avoid sharp corners and notches
Edge chipping / dullingMicro-chips along cutting edge; rapid loss of sharpnessImproper heat treatment (over-tempering, decarburization); excessive grinding heat (temper burn); coarse grit; impact loading beyond designVerify hardness HRC 48–53 per batch; control grinding temperature (no temper burn color); use fine grit finish on edge; design edge geometry for intended load; use 4Cr13 for maximum edge hardness

Failure mode descriptions are based on typical medical device field failure analysis. Root cause identification should always include metallographic examination of the failed part. Prevention measures should be validated for the specific device and documented in the design history file (DHF). For fatigue-critical components, finite element analysis (FEA) and fatigue testing per ISO 14801 (for orthopedic implants) or applicable device-specific standards are recommended.

Pitting corrosion is the most common failure in reusable surgical instruments, and it is almost always a surface contamination problem rather than a grade problem. When free iron from grinding wheels or carbon steel tooling becomes embedded in the stainless steel surface, it creates a galvanic cell that rusts during autoclaving, producing a rust spot that then pits into the base material. Passivation removes this free iron, but if passivation is skipped or degraded (old bath, insufficient time), the iron remains and pitting follows. The fix is process control: validated passivation baths, regular bath testing, and physical separation of stainless steel from carbon steel in the manufacturing area.

Fatigue failure is the most common failure in implantable and spring-loaded devices. It initiates at a surface defect — a grinding mark, a slit edge burr, a pit — and propagates under cyclic loading until the part fractures. Electropolishing is the single most effective prevention because it removes the surface defects where cracks start. For device springs made from SUS301, the full-hard condition (HV 430–520) provides the high tensile strength needed for fatigue resistance, but the surface must be free of scratches and burrs from the slitting and forming operations.

7. Grade Selection Summary

Work through these four questions to identify the correct grade for your medical device component.

  1. Does the component cut or need a hard, wear-resistant edge? If yes, specify 4Cr13 or SUS420J2 (martensitic, quench-hardened to HRC 48–53). 4Cr13 for maximum edge hardness; SUS420J2 for slightly better toughness and wider standard availability.
  2. Is the component implantable or in long-term contact with body tissue? If yes, specify SUS316L (ASTM F138/F139). Its low carbon and molybdenum addition provide the best corrosion resistance and biocompatibility for permanent implants.
  3. Does the component need to be a spring, clip or high-strength non-cutting part? If yes, specify SUS301 in the work-hardened condition (1/2-hard, 3/4-hard or full-hard). It combines austenitic corrosion resistance with high strength from cold working, without requiring heat treatment.
  4. Is the component a general housing, tray or non-critical structure? If yes, SUS304 is the standard and most economical choice. It offers good corrosion resistance for non-implant, non-cutting applications and is widely available in all gauges and finishes.

HS-FINEB supplies medical-grade stainless steel strip in 4Cr13, SUS420J2, SUS301, SUS304 and SUS316L, in thicknesses from 0.05 to 3.00 mm and widths from 10 to 600 mm, with precision slitting and edge conditioning processed in-house. Each coil ships with a mill test certificate reporting full chemistry and mechanical properties, and material is available with compliance documentation for ASTM F899, F138/F139 and ISO 7153-1 where applicable. For the full product range, see the stainless steel strip product page and the martensitic stainless steel strip page.

How HS-FINEB fits in

Medical-grade stainless steel strip — certified, traceable, precision-slit

HS-FINEB supplies stainless steel strip for medical device manufacturing across the full grade range — 4Cr13, SUS420J2, SUS301, SUS304 and SUS316L — with precision slitting, edge conditioning and cut-to-length processed at our Shanghai plant. Each coil ships with a mill test certificate reporting full chemistry and mechanical properties, with material traceability from heat to finished coil. Our strip is produced with controlled surface cleanliness to support downstream passivation and electropolishing, and we can provide compliance documentation for ASTM F899, F138/F139 and ISO 7153-1 where required. ISO 9001 certified. Send your grade, thickness, width, surface finish and annual volume for a quote within one working day.

Get a medical stainless steel quote →

Buyer FAQ

Medical stainless steel, asked and answered

Which stainless steel grade is best for surgical instruments?
For cutting surgical instruments (scalpels, scissors, osteotomes), martensitic grades such as 4Cr13 and SUS420J2 are the standard choice because they can be quenched and tempered to HRC 48–53, producing a hard, sharp edge that resists wear. For non-cutting instruments (forceps, retractors, trays), austenitic grades such as SUS304 and SUS316L are preferred because they offer superior corrosion resistance and cannot rust even after repeated sterilization, although they cannot be through-hardened. The grade choice depends on whether the instrument needs a hard cutting edge or maximum corrosion resistance.
Can SUS316L be hardened for surgical instrument use?
No. SUS316L is an austenitic stainless steel and cannot be hardened by heat treatment (quenching and tempering). Its as-supplied hardness is typically below HV 200 (HRC ~15 or lower). SUS316L can be work-hardened by cold working (cold rolling, drawing) to increase strength, but this is not the same as through-hardening and does not produce a hard, wear-resistant cutting edge. For instruments that need a hard edge, a martensitic grade such as SUS420J2 or 4Cr13 is required. SUS316L is used for implants, instrument housings, trays and components where corrosion resistance and biocompatibility matter more than edge hardness.
How many autoclave cycles can SUS420J2 survive before corroding?
Properly heat-treated and passivated SUS420J2 can withstand at least 1,000 cycles of 134°C steam autoclave sterilization without significant corrosion, provided the surface is free of embedded iron contamination and the passivation layer is intact. The actual number depends on surface finish (electropolished surfaces last longer than ground surfaces), water quality (chloride-free distilled water extends life), and whether the instruments are dried immediately after each cycle. Instruments showing rust spots after fewer than 100 cycles usually have a passivation failure or embedded free iron from manufacturing, not a grade problem.
What is the difference between 4Cr13 and SUS420J2?
4Cr13 (GB/T 1220) and SUS420J2 (JIS G4304) are closely related martensitic stainless steels with overlapping chemistry. 4Cr13 specifies C 0.36–0.45% and Cr 12.0–14.0%, while SUS420J2 specifies C 0.26–0.40% and Cr 12.0–14.0%. Both can be quenched to HRC 48–53 and are used for surgical blades and cutting instruments. 4Cr13 tends to run at the higher end of the carbon range, which can produce a slightly harder edge but with slightly lower toughness. They are often interchangeable, but for regulated medical devices, the grade specified on the drawing and the material certificate must match — confirm the exact standard and chemistry at RFQ.
Is electropolishing necessary for medical device stainless steel?
Electropolishing is not universally required, but it is strongly recommended for implantable devices and instruments that undergo repeated steam sterilization. Electropolishing reduces surface roughness from a typical Ra 0.4–0.8 μm (after grinding) to Ra ≤0.2 μm, removes the smeared surface layer and embedded free iron, and improves the uniformity of the passive chromium oxide layer. The result is better corrosion resistance, easier cleaning (fewer crevices for biofilm), and reduced risk of pitting during autoclaving. For non-implant, non-critical instruments, standard passivation may be sufficient.
Which sterilization method is most corrosive to stainless steel?
Steam autoclaving at 134°C is the most corrosive of the common sterilization methods because high-temperature, high-pressure moisture accelerates electrochemical corrosion, particularly if the water contains chloride ions or if the surface has embedded free iron. Ethylene oxide (EtO) sterilization is the least corrosive because it operates at low temperature (37–63°C) in a dry gas environment. Gamma radiation sterilization is also low-corrosion but can cause slight surface discoloration on some grades. For martensitic grades (SUS420J2, 4Cr13), proper passivation and drying after each autoclave cycle are essential to prevent staining and pitting.
More resources

Keep reading

Guides that follow from the same decision chain — martensitic stainless steel, surface finishes, edge burr control and mill test certificates.

Related steel families

Explore related product families

These grade families share material, processing or application territory with medical device stainless steel.

martensitic stainless steel strip for cutting instruments

4Cr13 / SUS420J2 / 440C grades quenched to HRC 48–58 for surgical blades and dental instruments.

SUS301 stainless steel strip for medical springs

Work-hardened austenitic strip up to HV 520 for device springs, clips and staples.

stainless steel strip for medical and precision use

Full range of austenitic and martensitic grades in precision gauges from 0.05 to 3.00 mm.

Need medical-grade stainless steel strip with certified chemistry?

Include grade, thickness × width, surface finish, standard (ASTM / JIS / GB) and annual volume. Our engineers reply within one working day.

Request a quote → View Stainless Steel Strip